Phoenix FA MCR-HT-1TS-I-OLP-top mounted transmitter 1145210, the front transmitter powered by the output loop can transmit RTD and TC sensor signals as well as signals from resistance sensors and voltage sensors through 4 mA ... 20 mA. HART, standard configuration, screw connection

Product Type temperature measurement transmitter
Product Series Field Analog
Channels 1
configuration Software
HART
insulation characteristics
Overvoltage Category II
pollution level 2

Function
configuration Software
HART

step response (0-99%) 1 s (TC)
1 s (RTD)
electrically isolated input/output
Test Voltage 2 kV AC (50 Hz, 60 s)
power supply
Supply Voltage Range loop power supply, no external power supply required
10V DC ... 36V DC (Standard product)

measurement
Input Number 1
Input Voltage Range -20 mV ... 100 mV (minimum measuring range: 5 mV)
Available Sensor Types (RTDs) Pt, Ni, Cu sensors: 2, 3, 4 wires
Available Sensor Types (TC) A, B, C, D, E, J, K, L, N, R, S, T, U
Temperature measurement range -250°C ... 2500°C (range depends on sensor type)
Temperature measurement range ≥ 10 K (RTD)
≥ 50 K (TC)
Linear resistance measuring range 10 Ω ... 2000 Ω (minimum measuring range: 10 Ω)

signals: current
current output signal 4 mA ... 20 mA (HART)
20 mA ... 4 mA (HART)
Maximum current output signal 23 mA
load (UL- 10 V) / 0.023 A
HART coding FSK ± 0.5 mA
HART protocol version 7
transmission speed 1200 Baud
Power Filter 50/60Hz
communication resistance ≥ 250 Ω
impact suppression  7 s (HART)
7 s (measured value)

Connection Mode screw connection
stripping length 8mm
thread M3
Rigid conductor cross section 0.2 mm² ... 2.5 mm²
Flexible Wire Cross Section 0.2 mm² ... 2.5 mm²
Attachable Guide Cross Section AWG 24 ... 14
Tightening torque 0.5 Nm

Explosion-proof installation (EPL) Gc
Div. 2

Data Communication (Bypass)
Supported Protocols HART

environmental conditions
protection level IP00 (IP66/67 installed in magnetic enclosure, NEMA 4X)
ambient temperature (operation) -40°C ... 85°C
ambient temperature (storage/transport) -50°C ... 100°C
elevation ≤ 4000 m (above sea level)
climate class C1
Allowable humidity (operating) 5%... 95% (condensable)
1145210

Phoenix FA MCR-HT-1TS-I-OLP-top mounted transmitter 1145210

Product Center

PRODUCT CENTER